Autor: |
Wolpert, Christian, Jung, Werner, Scholl, Christoph, Spehl, Susanne, Cyran, Joachim, Berndt Luderitz |
Předmět: |
|
Zdroj: |
Journal of Cardiovascular Electrophysiology; Aug1998, Vol. 9 Issue 8, p859-863, 5p |
Abstrakt: |
This case report describes delivery of atrial therapies during a sinus tachycardia in a new dual chamber implantable cardioverter defibrillator inappropriately caused by far-field oversensing of ventricular beats in the atrial channel. Upon classification of the PR interval pattern, the rate criterion for an atrial tachycardia was fulfilled, and the device initiated high-frequency burst pacing as the first stage of programmed tiered atrial therapies. Atrial fibrillation subsequently was induced by high-frequency burst pacing, and eventually a programmed 10-J shock was delivered for successful termination of atrial fibrillation. The phenomenon of far-field oversensing of ventricular beats could be repeatedly observed during exercise testing and abolished by decreasing the atrial sensitivity. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |