The effect of cochlear implant insertion technique on post-operative neural response telemetry and impedance in paediatric patients.

Autor: Hashemi, S B, Janipour, M, Jahangiri, R, Babaei, A
Předmět:
Zdroj: Journal of Laryngology & Otology; Apr2023, Vol. 137 Issue 4, p363-367, 5p
Abstrakt: Objective: This study aimed to compare neural response telemetry and impedance between the round window and cochleostomy approaches for cochlear implantation. Methods: In this case–control study, 64 patients aged less than 3.5 years underwent cochlear implantation via the round window or cochleostomy approach. Post-operative neural response telemetry and impedance were measured. Results: The impedance measurements at electrodes 1, 11 and 22 showed no significant differences between the two groups three months after implantation (p = 0.90, p = 0.08 and p = 0.37, respectively). Similar results were observed six months after implantation (p = 0.71, p = 0.65 and p = 0.70, respectively). There was no significant difference in neural response telemetry between the two groups after three months. The neural response telemetry of electrode 1 in the cochleostomy group (171.26 ± 19.81 μV) was significantly higher in comparison with that of electrode 1 in the round window group (161.97 ± 12.71 μV) after six months (p = 0.03). The neural response telemetry values for electrodes 11 and 22 did not show any significant difference after six months (p = 0.14 and p = 0.48, respectively). Conclusion: Both approaches provide equal stimulation of the cochlear nerve and impedance. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index