Autor: |
Sudarikov, I. S., Mindubaev, E. A. |
Předmět: |
|
Zdroj: |
Biomedical Engineering; Jul2022, Vol. 56 Issue 2, p100-104, 5p |
Abstrakt: |
Systems for transcutaneous inductive power transfer based on class E power amplifiers with series and parallel topologies for compensation of reactive power in the receiving part of the system were compared. The system with parallel compensation at the resonance frequency was shown to be a constant current source. In turn, in systems with series compensation, two possible frequencies can be selected, above and below the resonant frequency, at which the system is a constant voltage source. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|