Modeling the Dose Distribution in a Bone Implant during Sterilization with a Fast Electron Flow.

Autor: Khutsistova, A. O., Rozanov, V. V., Matveychuk, I. V., Chernyaev, A. P.
Zdroj: Moscow University Physics Bulletin; Dec2021 Supplement 1, Vol. 76 Issue 1, pS62-S67, 6p
Abstrakt: The passage of fast electron beams with energies in the range from 0.5 to 3 MeV through a fragment of bone tissue during radiation sterilization has been modeled using the GEANT4 software package. Analysis of the data showed that the optimal particle energies are 1 MeV for a sample thickness of 4 mm, 1.5 MeV for a sample thickness of 6 mm, 2 MeV for a sample thickness of 8 mm, and 2.5 MeV for a sample thickness of 10 mm when the sample is irradiated bilaterally or rotated. [ABSTRACT FROM AUTHOR]