Microgap and microleakage of a hybrid connection platform-switched implant system in the absence or presence of a silicone-based sealing agent.

Autor: Fernandes, Pedro Ferrás, Grenho, Liliana, Fernandes, Maria Helena, Sampaio-Fernandes, João Carlos, Sousa Gomes, Pedro
Předmět:
Zdroj: Odontology; Apr2022, Vol. 110 Issue 2, p231-239, 9p
Abstrakt: The present study aims to characterize, for the first time, the microgap and bacterial microleakage of a platform-switched implant system with hybrid connection, screwed at distinct torque values (manufacturer recommended torque—25 N cm—and a reduced torque—5 N cm—mimicking the long-term functional use), in the absence or presence of a silicon-based sealing agent. Microgap was determined through scanning electron microscopy and bacterial microleakage was evaluated in vitro, upon Enterococcus faecalis colonization of the system. The sealing efficacy was evaluated in the absence or presence of a commercially available silicon-based sealer. The cytotoxicity of the sealer was further addressed in vitro, with a fibroblastic cell line, in accordance with reference standards. A low microgap of the implant system was verified, regardless of the applied torque load—maximal values ranged around 0.25 and 1.25 µm, for 25 and 5 N cm torques, respectively. No bacterial microleakage was reported at 25 N cm, while at 5 N cm, leakage was verified on 38% of the samples. The application of a silicon-based sealer—with an adequate cytocompatible profile—was effective on preventing the bacterial microleakage on the assayed experimental setting. The assayed platform-switched implant system with hybrid connection presented a low interfacial misfit and an effective sealing capability at manufacturer recommended torque. Despite the increased microleakage at low torque conditions, the application of a cytocompatible silicon-based sealing agent restored the sealing effectiveness of the system. The use of a silicon-based sealing agent can assist on the maintenance of the sealing effectiveness even at low torque conditions. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index