Autor: |
Barbaraci, Gabriele |
Předmět: |
|
Zdroj: |
Journal of Vibration & Control; Dec2021, Vol. 27 Issue 23/24, p2711-2723, 13p |
Abstrakt: |
A mathematical discussion is introduced to describe the receiver coil characterizing a nuclear magnetic resonance for imaging, starting from a general shape of the conductor. A set of different inductance calculations have been introduced, varying the shape of the conductor. The inductance calculation led to a general expression of the magnetic field of a single coil characterized by a rectangular shape. A dynamic model of the receiver coil has been developed to represent the natural frequencies that characterize the operational bandwidth. A nonstationary control strategy is implemented to make a real time changing of the operational bandwidth. The frequency response of the coil generates the necessary conditional expression to let the peak of resonance move to a desired value of frequency. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |