A Portable Device for Rapid Assessment of the Functional State of the Hemostasis System.

Autor: Zhukov, E. L., Gubarev, F. A., Tyutrin, I. I.
Předmět:
Zdroj: Biomedical Engineering; Jan2021, Vol. 54 Issue 5, p345-349, 5p
Abstrakt: We present here the results obtained during the development of a portable device for rapid assessment of the functional state of the homeostasis system based on analysis of capillary blood. Vascular-thrombocytic hemostasis was assessed in volunteers in real time using the device. The results of piezocoagulography conducted using the portable device were compared to the results obtained by low-frequency piezothromboelastography (Mednord ARP-01M). These data showed a strong correlation, evidencing the potential of the method and device developed here for studies of the functional state of the hemostasis system at the patient's bedside or outside hospital. [ABSTRACT FROM AUTHOR]