Precision Measurements of Internal Conversion Coefficients of Low Energy Transitions in Tm for Efficiency Calibration of Electron Detectors.

Autor: Sai, K. Vijay, Rao, K. Madhusudhana, Rajasekhar, E., Rao, Dwaraka Rani, Seetharaman, Deepa, Kamisetti, Venkataramaniah
Předmět:
Zdroj: Physics of Atomic Nuclei; Nov2020, Vol. 83 Issue 6, p796-801, 6p
Abstrakt: The 32.018 d beta decay of Yb is studied with a high resolution precisely calibrated 8 PC based Multi-Channel Analyzer coupled HPGe gamma spectrometer system and liquid nitrogen cooled Si(Li) detector coupled to a well calibrated Mini-Orange magnetic spectrometer. Precise energies and relative intensities of gamma transitions and conversion electron intensities of gamma transitions have been determined for the first time with better accuracy. The experimental internal conversion coefficients of the gamma transitions in Tm have been determined using Normalized Peak to Gamma method and compared with the theoretical values adopted by Nuclear Data Sheets to assign multipolarities of all the transitions. The precise internal conversion electron intensities of the low energy transitions can be used as standards for calibration purposes of electron detectors and spectrometers. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index