Image plates as x-ray detectors in plasma physics experiments.

Autor: Gales, S. G., Bentley, C. D.
Předmět:
Zdroj: Review of Scientific Instruments; Oct2004 Part I & II, Vol. 75 Issue 10, p4001-4003, 3p
Abstrakt: The performance of image plates based on the photostimulable phosphor BaF(Br,l):Eu2+ has been investigated and compared with x-ray film. Evaluation of detective quantum efficiency (DQE), sensitivity, dynamic range, and linearity was carried out for several types of commercially available image plate, using the Excalibur soft x-ray calibration facility at AWE. Image plate response was found to be linear over a dynamic range of 5 orders of magnitude. One type of image plate was found to have a number of advantages for soft x-ray detection, with a measured sensitivity 1 order of magnitude greater than that of Kodak Industrex CX and DEF-5 x-ray film. The DQE of this plate was found to be superior to that of film at low [less than 103 photons/(50 μm)2] and high fluxes [greater than 104 photons/(50 μm)2]. The spatial resolution of image plates, scanned with several models of commercial image plate readers, has been evaluated using a USAF resolution test target. The highest spatial resolution measured is 35 μm. Though this is significantly lower than the resolution possible with film, it is sufficient for many applications. Image plates were fielded in a refractive x-ray lens imaging diagnostic on the 1 TW Helen laser and these results are discussed. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index