Autor: |
Brui, Ekaterina, Lepekhina, Anna, Chegina, Daria, Bukkieva, Tatiana, Efimtcev, Aleksandr, Andreychenko, Anna, Belov, Pavel, Petrov, Mihail |
Předmět: |
|
Zdroj: |
AIP Conference Proceedings; 2020, Vol. 2300 Issue 1, p1-4, 4p |
Abstrakt: |
This work is aimed to demonstrate the benefits of a new metamaterial-based radiofrequency coil for clinical 1.5 T hand and wrist MRI. Quantitative characteristics, such as signal-to-noise ratio and field-of-view, and the visual quality of MR images obtained with a standard clinical protocol for wrist imaging are compared between a conventional imaging setup (transmit-receive extremity coil) and a combination of a "birdcage" body coil with an inductively coupled split-loop- based coil. The study was conducted in a representative sample of subjects having different body masses and causing different coil loading. The utilization of the metamaterial-based coil provided large sensitive volume (field-of-view) - for 33% wider than for extremity coil, and allowed to gain the signal-to-noise ratio of all obtained images in a range from 1.26 to 1.40 times. The visual quality of all images acquired with the wireless coil was comparable with those acquired with the extremity coil. [ABSTRACT FROM AUTHOR] |