Autor: |
Chauhan, Vikas Singh, Bhushan, Sumit, Easwaran, Raghavan K. |
Předmět: |
|
Zdroj: |
AIP Conference Proceedings; 2020, Vol. 2241 Issue 1, p020012-1-020012-4, 4p |
Abstrakt: |
We have constructed in our laboratory two rectangular shaped current carrying coils in accordance with the parameters used in our design so that they can be put in use in the Two Dimensional Magneto Optical Trap (2D-MOT)to obtain very high value of Optical Depth(OD). The dimensions of the coils are 8cm x 10cm. When put in anti-Helmholtz configuration as required in a 2D-MOT, these coils generate magnetic field such that it becomes zero along a line bisecting the space between the two coils. The magnetic field gradient near the zero magnetic field line is measured to be 3.33 Gauss/cm when the current passing through the coils is 3.5A. We have also made measurements at other values of current and found, as expected, this gradient increases with current. These results match nicely with the estimated values and hence a significant step forward towards the experimental realization of the design proposed by us. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|