On Low-Energy Tail Distortions in the Detector Response Function of X-Ray Microcalorimeter Spectrometers.

Autor: O'Neil, G. C., Szypryt, P., Takacs, E., Tan, J. N., Buechele, S. W., Naing, A. S., Joe, Y. I., Swetz, D., Schmidt, D. R., Doriese, W. B., Gard, J. D., Reintsema, C. D., Ullom, J. N., Villarrubia, J. S., Ralchenko, Yu.
Předmět:
Zdroj: Journal of Low Temperature Physics; May2020, Vol. 199 Issue 3/4, p1046-1054, 9p
Abstrakt: We use narrow spectral lines from the X-ray spectra of various highly charged ions to measure low-energy tail-like deviations from a Gaussian response function in a microcalorimeter X-ray spectrometer with Au absorbers at energies from 650 to 3320 eV. We review the literature on low-energy tails in other microcalorimeter X-ray spectrometers and present a model that explains all the reviewed tail fraction measurements. In this model, a low-energy tail arises from the combination of electron escape and energy trapping associated with Bi X-ray absorbers. [ABSTRACT FROM AUTHOR]