Faster Determination of Mass-Averaged SAR From 2-D Area Scans.

Autor: Kanda, Michael Y., Douglas, Mark G., Mendivil, Edwin D., Ballen, Maurice, Gessner, Andrew V., Chou, Chung-Kwang
Předmět:
Zdroj: IEEE Transactions on Microwave Theory & Techniques; Aug2004 part 2 of 2, Vol. 52 Issue 8, p2013-2020, 8p
Abstrakt: An algorithm is presented that accurately and quickly estimates the peak 1- or beg averaged specific absorption rate (SAR) in a human phantom when exposed to a wireless device. Instead of performing both an area scan and a zoom scan (as per international standards), only the area scan and knowledge of the transmit frequency are needed. The accuracy of the algorithm has been demonstrated across a broad frequency range (150-2450 MHz) and for both 1- and 10-g averaged SAR using a sample of 264 SAR measurements from 55 wireless handsets. For the sample size studied, the root mean-squared errors of the algorithm are 1.2% and 5.8% for 1- and 10-g averaged SAR, respectively. It is shown that the algorithm works well in both head and body phantoms. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index