Evaluation of breath-hold 201Tl SPECT in the differential diagnosis of solitary pulmonary nodules.

Autor: Komori, Tsuyoshi, Narabayashi, Isamu, Hayashi, Masuo, Horiuchi, Shoji, Adachi, Itaru, Ogura, Yasuharu, Ohta, Hitoya, Utsunomiya, Keita
Zdroj: Annals of Nuclear Medicine; Jun2005, Vol. 19 Issue 4, p277-281, 5p
Abstrakt: Unlabelled: The aim of this study was to evaluate the usefulness of deep inspiration breath-hold SPECT (BrST, a method for 201Tl SPECT) in the diagnosis of solitary pulmonary nodules (SPN).Methods: Ten patients with malignant lesions and five with benign lesions were enrolled in this study. Early SPECT acquisition was performed 15 min after injection of 201Tl, while delayed SPECT images were acquired 3 h after injection. The first 15-sec acquisition was done using the BrST technique, and the second with the conventional free breathing (FB) method. We performed this technique alternately, and therefore, the odd data were from BrST and the even data were from FB. We referred to the T/N ratio of the early images as the ER and to the T/N ratio of the delayed images as the DR. To semi-quantitatively evaluate the degree of retention in the lesion, the retention index (RI) was calculated.Results: The RI of BrST indicated greater accuracy than that of FB in the differential diagnosis of SPN. For the benign and malignant lesions, the RI of BrST was -3.07 +/- 31.51 and 29.86 +/- 25.01, respectively (p < 0.05). The sensitivity, specificity, and accuracy of BrST (80%, 80%, and 80%, respectively) were significantly higher than those of FB (p < 0.05).Conclusion: The BrST method is more accurate than that of the conventional FB method in the differential diagnosis of SPN. [ABSTRACT FROM AUTHOR]