Autor: |
Pereira, J S, Pereira, M F, Rangel, S, Caldeira, M, Carvalhal, G, Santos, L M, Cardoso, J V, Alves, J G |
Předmět: |
|
Zdroj: |
Radiation Protection Dosimetry; Aug2019, Vol. 184 Issue 2, p216-223, 8p |
Abstrakt: |
In this work, the initial results of the type testing of the LiF:Mg,Cu,P (TLD-100H) whole-body personal dosemeters are presented. An assessment of reproducibility, linearity of the response, the residual signal as a function of the dose, energy and angular dependence of the response was performed. In general, the dosemeters show good reproducibility for different dose values and a linear behaviour for a range between 0.1 and 300 mSv. The detection limits obtained are lower than 50 μSv. The system presents a good energy and angular response for different radiation qualities. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|