Comparison of bolus- and filtering-based EIT measures of lung perfusion in an animal model.

Autor: Symon Stowe, Alistair Boyle, Michaël Sage, Wendy See, Jean-Paul Praud, Étienne Fortin-Pellerin, Andy Adler
Předmět:
Zdroj: Physiological Measurement; May2019, Vol. 40 Issue 5, p1-1, 1p
Abstrakt: Objective: Two main functional imaging approaches have been used to measure regional lung perfusion using electrical impedance tomography (EIT): venous injection of a hypertonic saline contrast agent and imaging of its passage through the heart and lungs, and digital filtering of heart-frequency impedance changes over sequences of EIT images. This paper systematically compares filtering-based perfusion estimates and bolus injection methods to determine to which degree they are related. Approach: EIT data was recorded on seven mechanically ventilated newborn lambs in which ventilation distribution was varied through changes in posture between prone, supine, left- and right-lateral positions. Perfusion images were calculated using frequency filtering and ensemble averaging during both ventilation and apnoea time segments for each posture to compare against contrast agent-based methods using Jaccard distance score. Main results: Using bolus-based EIT measures of lung perfusion as the reference frequency filtering techniques performed better than ensemble averaging and both techniques performed equally well across apnoea and ventilation data segments. Significance: Our results indicate the potential for use of filtering-based EIT measures of heart-frequency activity as a non-invasive proxy for contrast agent injection-based measures of lung perfusion. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index