Cryogen‐free dissolution dynamic nuclear polarization polarizer operating at 3.35 T, 6.70 T, and 10.1 T.

Autor: Ardenkjær‐Larsen, Jan Henrik, Bowen, Sean, Petersen, Jan Raagaard, Rybalko, Oleksandr, Vinding, Mads Sloth, Ullisch, Marcus, Nielsen, Niels Chr.
Zdroj: Magnetic Resonance in Medicine; Mar2019, Vol. 81 Issue 3, p2184-2194, 11p
Abstrakt: Purpose: A novel dissolution dynamic nuclear polarization (dDNP) polarizer platform is presented. The polarizer meets a number of key requirements for in vitro, preclinical, and clinical applications. Method: It uses no liquid cryogens, operates in continuous mode, accommodates a wide range of sample sizes up to and including those required for human studies, and is fully automated. Results: It offers a wide operational window both in terms of magnetic field, up to 10.1 T, and temperature, from room temperature down to 1.3 K. The polarizer delivers a 13C liquid state polarization for [1‐13C]pyruvate of 70%. The build‐up time constant in the solid state is approximately 1200 s (20 minutes), allowing a sample throughput of at least one sample per hour including sample loading and dissolution. Conclusion: We confirm the previously reported strong field dependence in the range 3.35 to 6.7 T, but see no further increase in polarization when increasing the magnetic field strength to 10.1 T for [1‐13C]pyruvate and trityl. Using a custom dry magnet, cold head and recondensing, closed‐cycle cooling system, combined with a modular DNP probe, and automation and fluid handling systems, we have designed a unique dDNP system with unrivalled flexibility and performance. [ABSTRACT FROM AUTHOR]