Readout of X-ray Pulses from a Single-pixel TES Microcalorimeter with Microwave Multiplexer Based on SQUIDs Directly Coupled to Resonators.

Autor: Nakashima, Y., Hirayama, F., Kohjiro, S., Yamamori, H., Nagasawa, S., Sato, A., Irimatsugawa, T., Muramatsu, H., Hayashi, T., Yamasaki, N. Y., Mitsuda, K.
Předmět:
Zdroj: Journal of Low Temperature Physics; Nov2018, Vol. 193 Issue 3/4, p618-625, 8p
Abstrakt: We have first demonstrated the cooperation between an X-ray TES and a directly coupled microwave SQUID multiplexer (D-coup. MW-Mux) in which SQUIDs are directly coupled to resonators. Using D-coup. MW-Mux, we successfully detected the X-ray pulses from a single-pixel TES and evaluated the energy resolution of 11.8 eV FWHM at 5.9 keV. We found this preliminary energy resolution is attributed to non-optimum bias point of the present TES, which can be improved by the future reduction of readout noise. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index
Nepřihlášeným uživatelům se plný text nezobrazuje