CALIBRATION PROCESS QUANTITY REDUCTION OF THE THERMAL VOLTAGE CONVERTER STANDARD USING A THREE-STAGE BUILD-UP AND BUILD-DOWN METHOD.

Autor: Sardjono, Hadi, Wijonarko, Sensus
Předmět:
Zdroj: International Journal of Technology; Jan2018, Vol. 9 Issue 1, p181-191, 11p
Abstrakt: Currently, three single junction-type Thermal Voltage Converter (TVC) standard units represent the highest standard of AC (Alternating Current) voltages owned by the Electrical Metrology Laboratory, Research Centre for Metrology--Indonesian Institute of Sciences. The accuracy of the single junction-type TVC is maintained regularly via intercomparison processes using a one-step build-up and build-down method. To reduce the calibration process quantity, three steps of build-up and build-down measurements that refer to the 4 V measurement point of a HOLT production single junction-type TVC were carried out. The dissemination processes with the best measurement accuracy up to 20 ppm were successfully obtained from measurement points between 1 V and 20 V via 4-1V, 4-2V, 4-3V, 4-6V, 4-10V, and 4-20V formations. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index