Giri, Upendra Kumar, Pradhan, Anirudh |
Předmět: |
|
Zdroj: |
Journal of Radiotherapy in Practice; Dec2017, Vol. 16 Issue 4, p409-414, 6p |
Abstrakt: |
ObjectiveThis study was conducted for the assessment of in-built systematic and random errors in the ExacTrac imaging system due to the software of Brainlab, on that basis; recommending a new quality control programme for ExacTrac imaging system.MethodsA program was developed to compare the image dataset of real time anthropomorphic pelvic phantom using ExacTrac with the reference image dataset from computed tomography. Images were acquired 20 times in a day, on single sitting for 20 conjugative days. On the basic of these translational and rotational shifts, systematic and random errors were calculated that had arisen due to multiple time image acquisition and image registration between acquired and reference image dataset of the phantom.ResultsRandom errors were found as 0·006 cm in right-left (Rt-Lt) direction, 0·008 cm in superior-inferior (Sup-Inf) direction and 0·012 cm in anterior-posterior (Ant-Post) direction. On this basic, margins were calculated using Van Herk formula; it was found that there were 0·02 cm inherent shift in Rt-Lt direction, 0·03 cm in Sup-Inf direction and 0·03 cm in Ant-Post direction.ConclusionThis study concluded that there was inherent error in ExacTrac system which can be quantified and used as a quality assurance tool for the ExacTrac system. [ABSTRACT FROM PUBLISHER] |