Effective luminance deterioration of medical liquid crystal displays in clinical use.

Autor: Takahashi, Keita, Awamoto, Shinichi, Takarabe, Shinya, Ogawa, Kazuhisa, Nakamura, Yasuhiko
Zdroj: Radiological Physics & Technology; Sep2017, Vol. 10 Issue 3, p382-386, 5p
Abstrakt: The objective of this work was to evaluate the maximum luminance ( L ) level of medical liquid crystal displays (LCDs) as a function of backlight hours (BLH) annually. The L values for 249 2-megapixel color LCDs (RadiForce RX210, EIZO Corporation) were measured in February 2014, 2015, and 2016. Four near-range luminance meters and the built-in type luminance meters, each with an LCD, were used for the measurements. The average and standard deviation (SD) of BLH measured in 2014 was 15,371 ± 8219 h. Four, twenty, and thirty-nine LCDs failed in the constancy tests performed in February 2014, 2015, and 2016, respectively, i.e., they were unable to output 170 cd/m. The SD of L increased each year and as BLH became longer. In conclusion, evaluation of L as a function of BLH during constancy testing will help predict the decrease in L of a clinically used medical color LCD. [ABSTRACT FROM AUTHOR]