Autor: |
Choi, Min, Kang, Gwansuk, Huh, Jung |
Zdroj: |
Biomedical Engineering Letters; May2017, Vol. 7 Issue 2, p143-151, 9p |
Abstrakt: |
This study was to optically visualize the cavitation bubbles produced by a clinical shock wave and to look into their geometric features of the resulting cavitation bubbles in relation to the driving shock wave field. A clinical shock wave therapeutic system was taken for shock wave production. The shock wave induced cavitation bubbles were captured by a professional camera under the illumination of a micro-pulse LED light. The light exposure was set to last for the whole life time of bubbles from formation to subsequent collapses. It was shown that the cavitation bubbles appeared mostly in the vicinity of the focus. The bubbles became more and larger as approaching to the focus. The cavitation bubbles formed jet streams which became enlarged (stronger) as the shock wave device output setting increased. The bubble cloud boundary was reasonably fitted to an elongated ellipsoid characteristically similar to the acoustic focal area. The bubble clouds were enlarged as the output setting increased. The geometric features of the cavitation bubbles characteristically similar to those of the focusing acoustic field have potential to provide the therapeutic focal information without time consuming hydrophone measurements of the shock wave field causing damages of the expensive sensor. The present study is limited to the static afterimages of the cavitation bubbles and investigation including the bubble dynamics is suggested to deliver the more realistic therapeutic area of the shock wave therapy [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|