Development of a synthetic single crystal diamond dosimeter for dose measurement of clinical proton beams.

Autor: Cyril Moignier, Dominique Tromson, Ludovic de Marzi, Fanny Marsolat, Juan Carlos García Hernández, Mathieu Agelou, Michal Pomorski, Romuald Woo, Jean-Michel Bourbotte, Fabien Moignau, Delphine Lazaro, Alejandro Mazal
Předmět:
Zdroj: Physics in Medicine & Biology; 7/7/2017, Vol. 62 Issue 13, p1-1, 1p
Abstrakt: The scope of this work was to develop a synthetic single crystal diamond dosimeter (SCDD-Pro) for accurate relative dose measurements of clinical proton beams in water. Monte Carlo simulations were carried out based on the MCNPX code in order to investigate and reduce the dose curve perturbation caused by the SCDD-Pro. In particular, various diamond thicknesses were simulated to evaluate the influence of the active volume thickness (eAV) as well as the influence of the addition of a front silver resin (250 µm in thickness in front of the diamond crystal) on depth-dose curves. The simulations indicated that the diamond crystal alone, with a small eAV of just 5 µm, already affects the dose at Bragg peak position (Bragg peak dose) by more than 2% with respect to the Bragg peak dose deposited in water. The optimal design that resulted from the Monte Carlo simulations consists of a diamond crystal of 1 mm in width and 150 µm in thickness with the front silver resin, enclosed by a water-equivalent packaging. This design leads to a deviation between the Bragg peak dose from the full detector modeling and the Bragg peak dose deposited in water of less than 1.2%. Based on those optimizations, an SCDD-Pro prototype was built and evaluated in broad passive scattering proton beams. The experimental evaluation led to probed SCDD-Pro repeatability, dose rate dependence and linearity, that were better than 0.2%, 0.4% (in the 1.0–5.5 Gy min−1 range) and 0.4% (for dose higher than 0.05 Gy), respectively. The depth-dose curves in the 90–160 MeV energy range, measured with the SCDD-Pro without applying any correction, were in good agreement with those measured using a commercial IBA PPC05 plane-parallel ionization chamber, differing by less than 1.6%. The experimental results confirmed that this SCDD-Pro is suitable for measurements with standard electrometers and that the depth-dose curve perturbation is negligible, with no energy dependence and no significant dose rate dependence. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index