Calibration of a Terahertz Attenuator by a DC Power Substitution Method.

Autor: Iida, Hitoshi, Kinoshita, Moto, Amemiya, Kuniaki
Předmět:
Zdroj: IEEE Transactions on Instrumentation & Measurement; Jun2017, Vol. 66 Issue 6, p1586-1591, 6p
Abstrakt: An attenuation calibration system is developed using an isothermal temperature control-type calorimeter for a terahertz (THz) attenuator. This system is based on a power ratio measurement by dc power substitution. A highly sensitive calorimeter achieves a simple and accurate attenuation calibration at room temperature. Calibration is performed for metalizedfilm attenuators at 1 THz using a photomixer as the THz source. The measurement uncertainties in the system are evaluated, and the expanded uncertainty is found to be 0.19–0.84 dB (k = 2) for an attenuation measurement range up to 12 dB. [ABSTRACT FROM AUTHOR]