Autor: |
Bockeria, L., Bockeria, O., Selishchev, S., Telyshev, D., Le, T., Satyukova, A., Shvartz, V., Glushko, L. |
Zdroj: |
Biomedical Engineering; Mar2017, Vol. 50 Issue 6, p416-419, 4p |
Abstrakt: |
The paper describes the procedure for evaluation of hemolysis induced by the Sputnik implantable pediatric rotary blood pump. In a series of five experiments the normalized index of hemolysis (N.I.H) for the Sputnik pump was found to be 0.0099 ± 0.0015 g/100 L. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|