Wearable Knee Health Rehabilitation Assessment using Acoustical Emissions.

Autor: Teague, Caitlin N., Hersek, Sinan, Conant, Jordan L., Gilliland, Scott M., Inan, Omer T.
Předmět:
Zdroj: AIP Conference Proceedings; 2017, Vol. 1806 Issue 1, p1-5, 5p, 2 Diagrams, 3 Graphs
Abstrakt: We have developed a novel, wearable sensing system based on miniature piezoelectric contact microphones formeasuring the acoustical emissions from the knee during movement. The system consists of two contactmicrophones, positioned on the medial and lateral sides of the patella, connected to custom, analog pre-amplifier circuits and a microcontroller for digitization and data storage on a secure digital card. In addition to the acoustical sensing, the system includes two integrated inertial measurement sensors including accelerometer and gyroscope modalities to enable joint angle calculations; these sensors, with digital outputs, are connected directly to the same microcontroller. The system provides low noise, accuratejoint acoustical emission and angle measurements in a wearable form factor and has several hours of battery life. [ABSTRACT FROM AUTHOR]