Autor: |
Bryant, Tony L., Anderson, Christopher L., Stevens, Christopher G., Conrad, Bryan P., Vincent, Heather K., Sadasivan, Kalia K. |
Zdroj: |
Journal of Orthopaedics; 2015, Vol. 12 Issue 2, p92-96, 5p |
Abstrakt: |
Purpose: This pilot study tested whether FiberWire provides similar protection to steel wire against repair displacement in patella fractures. Methods: Thirteen cadaver knees were cyclically loaded with 10 cycles (0-90° flexion) and fracture displacement was recorded. Fixation methods were also tested in load to failure (>3 mm displacement). Results: There was no difference between wire types in fracture displacement (1.4 mm ± 0.33 mm vs 1.2 mm ± 0.34 mm, respectively; p = 0.418) or in the load to failure (714.7 N ± 110.9 N vs 744.5 N ± 92.8 N, respectively; p = 0.360). Conclusion: FiberWire provides similar protection to steel wire against repair displacement after fixation of patellar fractures. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |
Externí odkaz: |
|