Cochlear implant users' spectral ripple resolution.

Autor: Eun Kyung Jeon, Turner, Christopher W., Karsten, Sue A., Henry, Belinda A., Gantz, Bruce J.
Předmět:
Zdroj: Journal of the Acoustical Society of America; Oct2015, Vol. 138 Issue 4, p2350-2358, 9p
Abstrakt: This study revisits the issue of the spectral ripple resolution abilities of cochlear implant (CI) users. The spectral ripple resolution of recently implanted CI recipients (implanted during the last 10 years) were compared to those of CI recipients implanted 15 to 20 years ago, as well as those of normal-hearing and hearing-impaired listeners from previously published data from Henry, Turner, and Behrens [J. Acoust. Soc. Am. 118, 1111-1121 (2005)]. More recently, implanted CI recipients showed significantly better spectral ripple resolution. There is no significant difference in spectral ripple resolution for these recently implanted subjects compared to hearing-impaired (acoustic) listeners. The more recently implanted CI users had significantly better pre-operative speech perception than previously reported CI users. These better pre-operative speech perception scores in CI users from the current study may be related to better performance on the spectral ripple discrimination task; however, other possible factors such as improvements in internal and external devices cannot be excluded. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index