Variations in depth-dose data between open and wedge fields for 4-MV X-rays.

Autor: Sewchand W, Khan FM, Williamson J
Jazyk: angličtina
Zdroj: Radiology [Radiology] 1978 Jun; Vol. 127 (3), pp. 789-92.
DOI: 10.1148/127.3.789
Abstrakt: Central-axis depth-dose data for 4-MV x rays, including tissue-maximum ratios, were measured for wedge fields. Comparison with corresponding open-field data revealed differences in magnitude which increased with depth, field size, and wedge thickness. However, phantom scatter correction factors for the wedge fields differed less than 1% from corresponding open-field factors. The differences in central-axis per cent depth doses between the two types of fields indicate beam hardening by the wedge filter. This study also implies that the derivation of tissue-maximum ratios from central-axis per cent depth is as valid for wedge as for open fields.