Long-term mechanical circulatory support system reliability recommendation: American Society for Artificial Internal Organs and Society of Thoracic Surgeons: long-term mechanical circulatory support system reliability recommendation.

Autor: Altieri F; ASAIO-STS Long-Term Circulatory Support System Reliability Working Group Recommendation., Berson A, Borovetz H, Butler K, Byrd G, Ciarkowski AA, Dunn R, Frazier OH, Griffith B, Hoeppner DW, Jassawalla JS, Kormos RH, Kung RT, Lemperle B, Lewis JP, Pantalos GM, Pennington DG, Poirier VL, Portner PM, Rosenberg G, Shanker R, Watson JT
Jazyk: angličtina
Zdroj: ASAIO journal (American Society for Artificial Internal Organs : 1992) [ASAIO J] 1998 Jan-Feb; Vol. 44 (1), pp. 108-14.
DOI: 10.1097/00002480-199801000-00020
Abstrakt: Jointly developed by members of the American Society for Artificial Internal Organs and the Society of Thoracic Surgeons along with staff from the Food and Drug Administration, the National Heart, Lung and Blood Institute and other experts, this recommendation describes the reliability considerations and goals for Investigational Device Exemption and Premarket Approval submissions for long-term, mechanical circulatory support systems. The recommendation includes a definition of system failure, a discussion of an appropriate reliability model, a suggested in vitro reliability test plan, reliability considerations for animal implantation tests, in vitro and animal in vivo performance goals, the qualification of design changes during the Investigational Device Exemption clinical trial, the development of a Failure Modes Effects and Criticality Analysis, and the reliability information for surgeons and patient candidates. The document will be periodically reviewed to assess its timeliness and appropriateness within five years.
Databáze: MEDLINE