Tandem reflectron time-of-flight mass spectrometer utilizing photodissociation.

Autor: Beussman DJ; Department of Chemistry, Michigan State University, East Lansing 48824, USA., Vlasak PR, McLane RD, Seeterlin MA, Enke CG
Jazyk: angličtina
Zdroj: Analytical chemistry [Anal Chem] 1995 Nov 01; Vol. 67 (21), pp. 3952-7.
DOI: 10.1021/ac00117a021
Abstrakt: A tandem time-of-flight (TOF) mass spectrometer has been designed to obtain complete MS/MS spectra from compounds eluting from a gas chromatograph. This application requires high spectral generation rate, unit mass resolution for both precursor selection and product spectra, and efficient ion utilization. These objectives are achieved by reflectron TOF mass separation in both stages and laser photoinduced dissociation as the ion fragmentation method. Careful timing of the laser pulse relative to ion extraction allows ions of a single m/z value up to m/z 1000 to be photodissociated while ions with adjacent m/z values are essentially unaffected. The convergent foci of the ion packet and laser pulse results in ion fragmentation efficiencies as high as 79%. An ion gate prevents the nonselected precursor ions from convoluting the product spectra. Product spectra can be generated at the maximum laser repetition rate (currently 200 Hz). To achieve unit mass resolution for all product m/z values simultaneously, a novel reflectron was designed for the second TOF stage.
Databáze: MEDLINE