MR imaging of the abdomen with a phased-array multicoil: prospective clinical evaluation.

Autor: Campeau NG; Department of Radiology, Mayo Clinic, Rochester, MN 55905, USA., Johnson CD, Felmlee JP, Rydberg JN, Butts RK, Ehman RL, Riederer SJ
Jazyk: angličtina
Zdroj: Radiology [Radiology] 1995 Jun; Vol. 195 (3), pp. 769-76.
DOI: 10.1148/radiology.195.3.7754009
Abstrakt: Purpose: To prospectively compare use of a phased-array multicoil and a conventional body coil in abdominal MR imaging.
Materials and Methods: Thirteen patients (seven men, six women; mean age, 55 years) underwent imaging with a phased-array multicoil and with a conventional body coil. Four pulse sequences were used: T2-weighted spin echo (SE), magnetization-prepared gradient-recalled echo (GRE), breath-hold fast SE, and echo planar (EP).
Results: Lesion detection improved the most on fast SE, multicoil-acquired images. Signal-to-noise ratio (S/N) increased 64% with fast SE (P = .0005) and EP (P < .0109) sequences. Contrast-to-noise ratio (C/N) doubled (P < .05) with T2-weighted SE sequences. Lesion conspicuity improved on multicoil-acquired images with all fast sequences (magnetization-prepared GRE, P = .015; fast SE, P = .002; EP imaging, P = .013). There was little difference in respiratory and vascular artifact. Depiction of most abdominal structures improved (P < .01).
Conclusion: Use of the phased-array multicoil provides better MR images of the abdomen than does use of a conventional body coil.