A technique of scatter and glare correction for videodensitometric studies in digital subtraction videoangiography.

Autor: Shaw CG, Ergun DL, Myerowitz PD, Van Lysel MS, Mistretta CA, Zarnstorff WC, Crummy AB
Jazyk: angličtina
Zdroj: Radiology [Radiology] 1982 Jan; Vol. 142 (1), pp. 209-13.
DOI: 10.1148/radiology.142.1.7031761
Abstrakt: The logarithmic amplification of video signals and the availability of data in digital form make digital subtraction videoangiography a suitable tool for videodensitometric estimation of physiological quantities. A system for this purpose was implemented with a digital video image processor. However, it was found that the radiation scattering and veiling glare present in the image-intensified video must be removed to make meaningful quantitations. An algorithm to make such a correction was developed and is presented. With this correction, the videodensitometry system was calibrated with phantoms and used to measure the left ventricular ejection fraction of a canine heart.