Autor: |
Vidiukov VI, Mikerova TM, Poliakov VN, Rotenberg EM, Dodel'tsev EP |
Jazyk: |
ruština |
Zdroj: |
Meditsinskaia radiologiia [Med Radiol (Mosk)] 1983 May; Vol. 28 (5), pp. 62-5. |
Abstrakt: |
The authors present a detailed description of the parameters of gamma-chambers that are necessary to measure in order to assess the quality of the device as a whole. The main concepts have been laid down: resolution, spatial distortion, irregularity of the field of vision of a gamma-chamber, idle time of the system, sensitivity, etc. Methods for the evaluation of the main characteristic of gamma-chambers are considered. A necessity to organize the checking of gamma-chambers in hospitals and other health care facilities and the making up of a common system of assessment are discussed. |
Databáze: |
MEDLINE |
Externí odkaz: |
|