The Reiter protein complement-fixation test using the Auto-Analyzer.

Autor: Pugh VW, Gaze RW
Jazyk: angličtina
Zdroj: Journal of clinical pathology [J Clin Pathol] 1966 Nov; Vol. 19 (6), pp. 595-9.
DOI: 10.1136/jcp.19.6.595
Abstrakt: A method for the Reiter protein complement-fixation test using the Techncon AutoAnalyzer is described. Two hundred and fifty sera were tested, when six false-positive and nine false-negative results were obtained when compared with results obtained by the manual method. Deep freezing the sera before testing on the AutoAnalyzer increased the sensitivity so that in a further 250 sera tested no false-negative and two false-positive results were obtained when compared with the results obtained at the Venereal Diseases Reference Laboratory, London.