Fronto-orbital advancement with patient-specific 3D-printed implants and robot-guided laser osteotomy: an in vitro accuracy assessment.
Autor: | Maintz M; Medical Additive Manufacturing Research Group (Swiss MAM), Department of Biomedical Engineering, University of Basel, Allschwil, Switzerland.; Institute for Medical Engineering and Medical Informatics, University of Applied Sciences and Arts Northwestern Switzerland, Muttenz, Switzerland.; Clinic Oral and Cranio-Maxillofacial Surgery, University Hospital Basel, Basel, Switzerland., Desan N; Medical Additive Manufacturing Research Group (Swiss MAM), Department of Biomedical Engineering, University of Basel, Allschwil, Switzerland., Sharma N; Medical Additive Manufacturing Research Group (Swiss MAM), Department of Biomedical Engineering, University of Basel, Allschwil, Switzerland. neha.sharma@unibas.ch.; Clinic Oral and Cranio-Maxillofacial Surgery, University Hospital Basel, Basel, Switzerland. neha.sharma@unibas.ch., Beinemann J; Clinic Oral and Cranio-Maxillofacial Surgery, University Hospital Basel, Basel, Switzerland., Beyer M; Medical Additive Manufacturing Research Group (Swiss MAM), Department of Biomedical Engineering, University of Basel, Allschwil, Switzerland.; Clinic Oral and Cranio-Maxillofacial Surgery, University Hospital Basel, Basel, Switzerland., Seiler D; Institute for Medical Engineering and Medical Informatics, University of Applied Sciences and Arts Northwestern Switzerland, Muttenz, Switzerland., Honigmann P; Medical Additive Manufacturing Research Group (Swiss MAM), Department of Biomedical Engineering, University of Basel, Allschwil, Switzerland.; Hand- and Peripheral Nerve Surgery, Department of Orthopaedic Surgery and Traumatology, Kantonsspital Baselland, Bruderholz Liestal Laufen, Switzerland.; Biomedical Engineering and Physics, Amsterdam UMC Location University of Amsterdam, Amsterdam, The Netherlands., Soleman J; Department of Neurosurgery, University Hospital Basel, Basel, Switzerland.; Division of Pediatric Neurosurgery, Children's University Hospital Basel, Basel, Switzerland.; Faculty of Medicine, University of Basel, Basel, Switzerland., Guzman R; Department of Neurosurgery, University Hospital Basel, Basel, Switzerland.; Division of Pediatric Neurosurgery, Children's University Hospital Basel, Basel, Switzerland.; Faculty of Medicine, University of Basel, Basel, Switzerland., Cattin PC; Center of Medical Image Analysis and Navigation (CIAN), Department of Biomedical Engineering, University of Basel, Allschwil, Basel, Switzerland., Thieringer FM; Medical Additive Manufacturing Research Group (Swiss MAM), Department of Biomedical Engineering, University of Basel, Allschwil, Switzerland.; Clinic Oral and Cranio-Maxillofacial Surgery, University Hospital Basel, Basel, Switzerland.; Faculty of Medicine, University of Basel, Basel, Switzerland. |
---|---|
Jazyk: | angličtina |
Zdroj: | International journal of computer assisted radiology and surgery [Int J Comput Assist Radiol Surg] 2024 Dec 13. Date of Electronic Publication: 2024 Dec 13. |
DOI: | 10.1007/s11548-024-03298-6 |
Abstrakt: | Purpose: The use of computer-assisted virtual surgical planning (VSP) for craniosynostosis surgery is gaining increasing implementation in the clinics. However, accurately transferring the preoperative planning data to the operating room remains challenging. We introduced and investigated a fully digital workflow to perform fronto-orbital advancement (FOA) surgery using 3D-printed patient-specific implants (PSIs) and cold-ablation robot-guided laser osteotomy. This novel approach eliminates the need for traditional surgical templates while enhancing precision and customization, offering a more streamlined and efficient surgical process. Methods: Computed tomography data of a patient with craniosynostosis were used to digitally reconstruct the skull and to perform VSP of the FOA. In total, six PSIs per skull were 3D-printed with a medical-grade bioresorbable composite using the Arburg Plastic Freeforming technology. The planned osteotomy paths and the screw holes, including their positions and axis angles, were digitally transferred to the cold-ablation robot-guided osteotome interface. The osteotomies were performed on 3D-printed patient skull models. The implants, osteotomy and final FOA results were scanned and compared to the VSP data. Results: The osteotomy deviations for the skulls indicated an overall maximum distance of 1.7 mm, a median deviation of 0.44 mm, and a maximum root mean square (RMS) error of 0.67 mm. The deviation of the point-to-point surface comparison of the FOA with the VSP data resulted in a median accuracy of 1.27 mm. Accessing the orbital cavity with the laser remained challenging. Conclusion: This in vitro study showcases a novel FOA technique by effectively combining robot-guided laser osteotomy with 3D-printed patient-specific implants, eliminating the need for surgical templates and achieving high accuracy in bone cutting and positioning. The workflow holds promise for reducing preoperative planning time and increasing surgical efficiency. Further studies on bone tissue are required to validate the safety and effectiveness of this approach, especially in addressing the challenges of pediatric craniofacial surgery. Competing Interests: Declarations. Conflict of interest: Philippe Cattin is a co-founder of AOT AG. The remaining authors declare no conflict of interest. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript, or in the decision to publish the results. (© 2024. The Author(s).) |
Databáze: | MEDLINE |
Externí odkaz: |