Retrospective analysis of treatment-positioning accuracy and dose error in boron neutron capture therapy using a sitting-position treatment system for head and neck cancer.

Autor: Komori S; Department of Radiation Physics and Technology, Southern Tohoku BNCT Research Center, 7-10 Yatsuyamada, Koriyama, Fukushima 963-8052, Japan. Electronic address: skomori.0922@gmail.com., Hirose K; Department of Radiation Oncology, Hirosaki University Graduate School of Medicine, 5 Zaifu-cho, Hirosaki, Aomori, 036-8562, Japan., Sato M; Department of Radiation Oncology, Hirosaki University Graduate School of Medicine, 5 Zaifu-cho, Hirosaki, Aomori, 036-8562, Japan., Yamazaki Y; Department of Radiation Physics and Technology, Southern Tohoku BNCT Research Center, 7-10 Yatsuyamada, Koriyama, Fukushima 963-8052, Japan., Takeuchi A; Department of Radiation Physics and Technology, Southern Tohoku BNCT Research Center, 7-10 Yatsuyamada, Koriyama, Fukushima 963-8052, Japan., Kato R; Department of Radiation Physics and Technology, Southern Tohoku BNCT Research Center, 7-10 Yatsuyamada, Koriyama, Fukushima 963-8052, Japan., Motoyanagi T; Department of Radiation Physics and Technology, Southern Tohoku BNCT Research Center, 7-10 Yatsuyamada, Koriyama, Fukushima 963-8052, Japan., Narita Y; Department of Radiation Physics and Technology, Southern Tohoku BNCT Research Center, 7-10 Yatsuyamada, Koriyama, Fukushima 963-8052, Japan., Kato T; School of Health Sciences, Fukushima Medical University, 10-6 Sakaemachi, Fukushima, Fukushima, 960-8516, Japan., Takai Y; Department of Radiation Oncology, Southern Tohoku BNCT Research Center, 7-10 Yatsuyamada, Koriyama, Fukushima, 963-8052, Japan.
Jazyk: angličtina
Zdroj: Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) [Phys Med] 2024 Oct; Vol. 126, pp. 104818. Date of Electronic Publication: 2024 Oct 01.
DOI: 10.1016/j.ejmp.2024.104818
Abstrakt: The neutron beam in boron neutron capture therapy (BNCT) exhibits poor directionality and significantly decreasing neutron flux with increasing distance. Therefore, the treatment site must be close to the irradiation aperture. Some patients with head and neck cancer may benefit from a sitting-position setup. The study aim was to evaluate the treatment-positioning accuracy and dose error in sitting patients receiving BNCT. Thirty-two patients with head and neck cancer who underwent sitting-position BNCT at Southern Tohoku BNCT Research Center were included in the study. Horizontal (ΔX) and vertical (ΔY) errors were defined as the displacement between the treatment planning system (TPS) digital reconstructed radiograph and the pre-treatment X-ray image. Using in-house software, image matching was performed. The beam-axial directional (ΔZ) error was compared with the parameters entered into the TPS and the actual pre-treatment measured values. The translational-position error was reflected in the TPS's patient coordinate system with respect to the reference plan. Re-dose calculations were performed to evaluate the effect of positional error on tumor and normal-tissue doses. The [ΔX, ΔY, ΔZ] DRR-CR mean ± 1SD were - 0.40 ± 2.0, 0.30 ± 2.3, and - 1.4 ± 1.5 mm, respectively. The D mean and D 98% tumor-dose errors were 1.22 % ± 1.44 % and 0.99 % ± 1.63 %, respectively. The D 2% pharyngeal and oral mucosal-dose errors were 0.98 % ± 1.91 % and 1.21 % ± 1.78 %, respectively. The tumor- and normal-tissue dose errors were typically < 5 %. High-precision treatment was feasible in sitting-positioned BNCT.
Competing Interests: Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
(Copyright © 2024 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved.)
Databáze: MEDLINE