Combining Enhanced Resolving Power with Duty Cycle Improvements on a Multi-Reflecting Time-of-Flight Mass Spectrometer.

Autor: Johnson WJ; Waters Corporation, Stamford Avenue, Altrincham Road, Wilmslow, Cheshire SK9 4AX, U.K., Palmer ME; Waters Corporation, Stamford Avenue, Altrincham Road, Wilmslow, Cheshire SK9 4AX, U.K., Claude E; Waters Corporation, Stamford Avenue, Altrincham Road, Wilmslow, Cheshire SK9 4AX, U.K., McCullagh M; Waters Corporation, Stamford Avenue, Altrincham Road, Wilmslow, Cheshire SK9 4AX, U.K., Nixon P; Waters Corporation, Stamford Avenue, Altrincham Road, Wilmslow, Cheshire SK9 4AX, U.K., Wildgoose J; Waters Corporation, Stamford Avenue, Altrincham Road, Wilmslow, Cheshire SK9 4AX, U.K.
Jazyk: angličtina
Zdroj: Journal of the American Society for Mass Spectrometry [J Am Soc Mass Spectrom] 2024 Sep 04; Vol. 35 (9), pp. 2073-2081. Date of Electronic Publication: 2024 Aug 09.
DOI: 10.1021/jasms.4c00122
Abstrakt: The combination of enhanced resolving power and improved duty cycle on a multireflecting time-of-flight mass spectrometer is described. Resolving power increases are achieved by extending the effective ion path length from 47 m to greater than 200 m. Path length increases are achieved through containment of ions within the analyzer for up to N = 5 passes using a pulsed deflection electrode. Resolving power was shown to increase from 220,000 to 402,000 (fwhm) at m / z 785 for N = 1 and N = 4 analyzer passes, respectively. Due to the timing of the pulsed deflection electrode, the approach is particularly suited to high resolution analysis over a targeted m / z range. Duty cycle enhancements are achieved for ions of the targeted m / z range via accumulation prior to orthogonal acceleration, providing signal improvements of 2 orders of magnitude. Achieving such high resolving powers at fast scan rates (30 Hz) can yield additional information such as fine isotope structure; when combined with ppb mass measurement accuracy, high confidence in analyte identification can be achieved. The technique is applied for N = 2 analyzer passes, demonstrating fine isotope structure for a typical UHPLC metabolite identification experiment at a 10 Hz acquisition rate. Additionally, mass spectrometry imaging data is acquired using DESI, demonstrating the improved image clarity achieved at >300,000 (fwhm).
Databáze: MEDLINE