Effect of a 3-second Off-label Exposure on the Depth of Cure of Eight Resin-based Composites.

Autor: Ribeiro M; Maria Tereza Hordones Ribeiro DDS, MSc, PhD student, Department of Operative Dentistry and Dental Materials, Dental School, Federal University of Uberlândia, Minas Gerais, Brazil., Maucoski C; Cristiane Maucoski, DDS, MSc, PhD, Department of Operative Dentistry and Dental Materials, Dental School, Federal University of Ponta Grossa, Ponta Grossa, Paraná, Brazil., Price RB; Richard Bengt Price, DDS, MS, PhD, professor, Department of Dental Clinical Sciences, Dalhousie University, Halifax, Nova Scotia, Canada., Soares CJ; Carlos José Soares, DDS, MSc, PhD, professor, Department of Operative Dentistry and Dental Materials, Biomechanics Research Center, Federal University of Uberlândia, Minas Gerais, Brazil.
Jazyk: angličtina
Zdroj: Operative dentistry [Oper Dent] 2024 Jul 01; Vol. 49 (4), pp. 421-431.
DOI: 10.2341/23-155-L
Abstrakt: Objectives: This study evaluated the depth of cure (DoC) of eight resin-based composites (RBCs) photocured using one multipeak light-curing unit (LCU) on the standard output setting for the manufacturer's RBC recommended exposure time and at a higher irradiance for 3 seconds.
Methods: Three conventional RBCs: Tetric EvoCeram (Evo), Tetric N-Ceram (Cer), Tetric Prime (Pri); and five bulk-fill: Tetric N-Ceram Bulk Fill (CerBF), Opus Bulk Fill APS (OpusBF), Opus Bulk Fill Flow APS (OpusF), Tetric PowerFill (PFill) and Tetric PowerFlow (PFlow) were examined. Only PFill and PFlow are formulated to be photocured in 3 seconds. The RBCs were packed into a metal mold and photocured using a Bluephase PowerCure LCU for the RBC manufacturer's recommended exposure time on the standard mode and using the 3-second high irradiance mode. After photocuring, the specimens were immersed in a solvent for 1 hour. The length of the remaining RBC was measured and divided by 2. Data were analyzed using two-way analysis of variance (ANOVA) followed by the Tukey post hoc multiple comparison test (α=0.05).
Results: There was no significant difference in the DoC values for PFill and PFlow when photocured using the 3-second high irradiance protocol compared to the lower irradiance standard mode protocol. All other RBCs had significantly lower DoC values (p<0.001) when photocured off-label using the 3-second high irradiance mode.
Conclusion: Of the eight RBCs tested, only PFill and PFlow achieved the same DoC when the high irradiance 3-second curing method was used compared to when their longer lower irradiance protocol was used.
(©Operative Dentistry, 2024.)