Evaluation and improvement of the safety of 3D-printed template assisted intracavitary/interstitial brachytherapy for cervical cancer using repeat FMEA.
Autor: | Shen J; Department of Radiation and Medical Oncology, Hubei Key Laboratory of Tumor Biological Behaviors, Hubei Cancer Clinical Study Center, Zhongnan Hospital of Wuhan University, Wuhan, China., Chen M; Department of Radiation and Medical Oncology, Hubei Key Laboratory of Tumor Biological Behaviors, Hubei Cancer Clinical Study Center, Zhongnan Hospital of Wuhan University, Wuhan, China., Qiu H; Department of Radiation and Medical Oncology, Hubei Key Laboratory of Tumor Biological Behaviors, Hubei Cancer Clinical Study Center, Zhongnan Hospital of Wuhan University, Wuhan, China., Yang C; Department of Radiation and Medical Oncology, Hubei Key Laboratory of Tumor Biological Behaviors, Hubei Cancer Clinical Study Center, Zhongnan Hospital of Wuhan University, Wuhan, China., Liu H; Department of Radiation and Medical Oncology, Hubei Key Laboratory of Tumor Biological Behaviors, Hubei Cancer Clinical Study Center, Zhongnan Hospital of Wuhan University, Wuhan, China., Chen J; Department of Radiation and Medical Oncology, Hubei Key Laboratory of Tumor Biological Behaviors, Hubei Cancer Clinical Study Center, Zhongnan Hospital of Wuhan University, Wuhan, China., Wang D; Department of Radiation and Medical Oncology, Hubei Key Laboratory of Tumor Biological Behaviors, Hubei Cancer Clinical Study Center, Zhongnan Hospital of Wuhan University, Wuhan, China., Zhao H; Department of Radiation and Medical Oncology, Hubei Key Laboratory of Tumor Biological Behaviors, Hubei Cancer Clinical Study Center, Zhongnan Hospital of Wuhan University, Wuhan, China., Sun S; Department of Radiation and Medical Oncology, Hubei Key Laboratory of Tumor Biological Behaviors, Hubei Cancer Clinical Study Center, Zhongnan Hospital of Wuhan University, Wuhan, China., Mei Z; Department of Radiation and Medical Oncology, Hubei Key Laboratory of Tumor Biological Behaviors, Hubei Cancer Clinical Study Center, Zhongnan Hospital of Wuhan University, Wuhan, China., Xie C; Department of Radiation and Medical Oncology, Hubei Key Laboratory of Tumor Biological Behaviors, Hubei Cancer Clinical Study Center, Zhongnan Hospital of Wuhan University, Wuhan, China., Wang X; Department of Radiation and Medical Oncology, Hubei Key Laboratory of Tumor Biological Behaviors, Hubei Cancer Clinical Study Center, Zhongnan Hospital of Wuhan University, Wuhan, China.. Electronic address: wangxiaoyong@znhospital.cn. |
---|---|
Jazyk: | angličtina |
Zdroj: | Brachytherapy [Brachytherapy] 2024 Sep-Oct; Vol. 23 (5), pp. 580-589. Date of Electronic Publication: 2024 Jul 02. |
DOI: | 10.1016/j.brachy.2024.05.003 |
Abstrakt: | Background and Purpose: 3D-printed templates are used in intracavitary/interstitial brachytherapy (3DP-IC/IS) for locally advanced cervical cancer (LACC). We applied failure mode and effects analysis (FMEA) twice in one year to improve 3DP-IC/IS safety. Materials and Methods: A risk assessment group was established. We created a process map for 3DP-IC/IS procedures, identifying potential failure modes (FMs) and evaluating occurrence (O), detectability (D), severity (S), and risk priority number (RPN = O*D*S). High RPN values identified high-risk FMs, and quality control (QC) methods were determined by root cause analysis. A second FMEA was performed a year later. Results: The 3DP-IC/IS process included 10 main steps, 48 subprocesses, and 54 FMs. Initial RPN values ranged from 4.50 to 171.00 (median 50.50; average 52.18). Ten high-risk FMs were identified: (1) unreasonable needle track design (171.00/85.50), (2) noncoplanar needle label identification failure (126.00/64.00), (3) template model reconstruction failure (121.50/62.50), (4) improper gauze filling (112.00/60.25), (5) poor needle position (112.00/52.50). QC interventions lowered all high-risk RPN values during the second assessment. Conclusions: A feasible 3DP-IC/IS process was proposed. Staff training, automatic needle path planning, insertion guidance diagrams, template checking, system commissioning, and template design improvements effectively enhanced process safety. (Copyright © 2024 The Authors. Published by Elsevier Inc. All rights reserved.) |
Databáze: | MEDLINE |
Externí odkaz: |