Dose to organs at risk for total body irradiation: Single-institution data using the modulated arc total body irradiation technique.

Autor: Effeney B; Radiation Oncology Princess Alexandra Hospital - Raymond Terrace, South Brisbane, Queensland, Australia., Pullar A; Radiation Oncology Princess Alexandra Hospital - Raymond Terrace, South Brisbane, Queensland, Australia., Burbery J; School of Clinical Sciences, Faculty of Health, Queensland University of Technology, Brisbane, Queensland, Australia., Hargrave C; Radiation Oncology Princess Alexandra Hospital - Raymond Terrace, South Brisbane, Queensland, Australia.; School of Clinical Sciences, Faculty of Health, Queensland University of Technology, Brisbane, Queensland, Australia., Brady C; Radiation Oncology Princess Alexandra Hospital - Raymond Terrace, South Brisbane, Queensland, Australia.
Jazyk: angličtina
Zdroj: Pediatric blood & cancer [Pediatr Blood Cancer] 2024 Sep; Vol. 71 (9), pp. e31164. Date of Electronic Publication: 2024 Jul 02.
DOI: 10.1002/pbc.31164
Abstrakt: Background: Organs at risk (OAR) dose reporting for total body irradiation (TBI) patients is limited, and standardly reported only as mean doses to the lungs and kidneys. Consequently, dose received and effects on other OAR remain unexplored. To remedy this gap, this study reports dose data on an extensive list of OAR for patients treated at a single institution using the modulated arc total body irradiation (MATBI) technique.
Method: An audit was undertaken of all patients treated with MATBI between January 2015 and March 2021 who had completed their course of treatment. OAR were contoured on MATBI patient treatment plans, with 12 Gy in six fraction prescription. OAR dose statistics and dose volume histogram data are reported for the whole body, lungs, kidneys, bones, brain, lens, heart, liver and bowel bag.
Results: The OAR dose data for 29 patients are reported. Mean dose results are body 11.77 Gy, lungs 9.86 Gy, kidneys 11.84 Gy, bones 12.03 Gy, brain 12.12 Gy, right lens 12.31 Gy, left lens 12.64 Gy, heart 11.07 Gy, liver 11.81 Gy and bowel bag 12.06 Gy. Dose statistics at 1-Gy intervals of V6-V13 for lungs and V10-V13 for kidneys are also included.
Conclusion: This is the first time an extensive list of OAR data has been reported for any TBI technique. Due to the paucity of reporting, this information could be used by centres implementing the MATBI technique, in addition to aiding comparison between TBI techniques, with the potential for greater understanding of the relationship between dose volume data and toxicity.
(© 2024 Wiley Periodicals LLC.)
Databáze: MEDLINE