Post cataract surgery refractive surprise due to intraocular lens mislabeling.

Autor: Shokoohi Rad S; Eye Research Center Mashhad University of Medical Sciences Mashhad Iran., Ansari Astaneh MR; Eye Research Center Mashhad University of Medical Sciences Mashhad Iran., Kiarudi MY; Eye Research Center Mashhad University of Medical Sciences Mashhad Iran., Ghavami Shahri SH; Eye Research Center Mashhad University of Medical Sciences Mashhad Iran., Heidarzadeh HR; Eye Research Center Mashhad University of Medical Sciences Mashhad Iran.
Jazyk: angličtina
Zdroj: Clinical case reports [Clin Case Rep] 2024 Jun 21; Vol. 12 (7), pp. e9104. Date of Electronic Publication: 2024 Jun 21 (Print Publication: 2024).
DOI: 10.1002/ccr3.9104
Abstrakt: Key Clinical Message: If an intraocular lens (IOL) is incorrectly labeled, problems can arise for surgeons and patients. Precise biometry is important to accurately determine the IOL power and prevent the implantation of the wrong IOL. Labeling and verifying IOLs with accuracy is crucial to ensuring the best possible results of cataract surgery.
Abstract: Mislabeling of IOLs can cause unpredictable problems for surgeons. However, we can prevent incorrect IOLs by using precise biometry to determine accurate IOL power and safely implant the correct IOL. A 50-year-old female with no medical or ocular history came to our clinic complaining of decreased vision in both eyes that had been ongoing for several months. After being diagnosed with cataracts, primary angle closure suspect, and high hyperopia, the patient underwent phacoemulsification surgery. A posterior chamber IOL was implanted, and visco-goniosynechialysis was performed. During follow-up appointments, it was discovered that the patient had an uncorrected visual acuity of 20/50 in her right eye, which was corrected to 20/20 with a + 7.00 D lens. Upon further evaluation, it was determined that the source of the error was due to a manufacturing mislabeling of the IOL power. The patient then underwent successful IOL exchange surgery, and her best-corrected visual acuity became 20/20 with no significant refraction. This case highlights an uncommon source of refractive surprise after phacoemulsification surgery, successfully managed with IOL exchange surgery using the same IOL power from a different brand.
Competing Interests: The authors declare that they have no competing interests.
(© 2024 The Author(s). Clinical Case Reports published by John Wiley & Sons Ltd.)
Databáze: MEDLINE
Nepřihlášeným uživatelům se plný text nezobrazuje