Relationship between oscillatory volume and oscillatory pressure in neonatal high-frequency oscillatory ventilators: an in vitro study.
Autor: | Zannin E; Neonatal Intensive Care Unit, Fondazione IRCCS San Gerardo dei Tintori, Monza, Italy., Rigotti C; Neonatal Intensive Care Unit, Fondazione IRCCS San Gerardo dei Tintori, Monza, Italy camilla.rigotti@irccs-sangerardo.it., Dellacà RL; Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Milano, Italy., Ventura ML; Neonatal Intensive Care Unit, Fondazione IRCCS San Gerardo dei Tintori, Monza, Italy. |
---|---|
Jazyk: | angličtina |
Zdroj: | Archives of disease in childhood [Arch Dis Child] 2024 Jul 18; Vol. 109 (8), pp. 666-672. Date of Electronic Publication: 2024 Jul 18. |
DOI: | 10.1136/archdischild-2023-326742 |
Abstrakt: | Objective: We analysed the relationship between oscillatory volume (V Design: In vitro study. Setting: Neonatal intensive care unit. Interventions: Ventilators tested were VN800 (Dräger), Servo-n (Maquet Getinge), SensorMedics 3100A (Vyaire Medical), Fabian HFOi (Vyaire Medical), SLE6000 (SLE UK) and Humming Vue (Metran). We changed various settings and mechanical characteristics of the test lung to mimic preterm and term conditions. Main Outcome Measures: For each condition, we measured V Results: We observed large variability between machines in the ΔP displayed at any target V Conclusions: The variability in the relationship between V Competing Interests: Competing interests: RLD reports that Politecnico di Milano received research grants from Vyaire. CR has received honoraria for lectures from Vyaire. (© Author(s) (or their employer(s)) 2024. No commercial re-use. See rights and permissions. Published by BMJ.) |
Databáze: | MEDLINE |
Externí odkaz: |