Auditory Effects of Acoustic Noise From 3-T Brain MRI in Neonates With Hearing Protection.

Autor: Jin C; Department of Radiology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, Shaanxi, China.; Shaanxi Engineering Research Center of Computational Imaging and Medical Intelligence, Xi'an, Shaanxi, China., Zhao H; Department of Radiology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, Shaanxi, China.; Shaanxi Engineering Research Center of Computational Imaging and Medical Intelligence, Xi'an, Shaanxi, China., Li H; Department of Radiology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, Shaanxi, China.; Shaanxi Engineering Research Center of Computational Imaging and Medical Intelligence, Xi'an, Shaanxi, China., Chen P; Department of Radiology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, Shaanxi, China.; Shaanxi Engineering Research Center of Computational Imaging and Medical Intelligence, Xi'an, Shaanxi, China., Tian C; Department of Radiology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, Shaanxi, China.; Shaanxi Engineering Research Center of Computational Imaging and Medical Intelligence, Xi'an, Shaanxi, China., Li X; Department of Radiology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, Shaanxi, China.; Shaanxi Engineering Research Center of Computational Imaging and Medical Intelligence, Xi'an, Shaanxi, China., Wang M; Department of Radiology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, Shaanxi, China.; Shaanxi Engineering Research Center of Computational Imaging and Medical Intelligence, Xi'an, Shaanxi, China., Liu C; Department of Radiology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, Shaanxi, China.; Shaanxi Engineering Research Center of Computational Imaging and Medical Intelligence, Xi'an, Shaanxi, China., Sun Q; Department of Radiology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, Shaanxi, China.; Shaanxi Engineering Research Center of Computational Imaging and Medical Intelligence, Xi'an, Shaanxi, China., Zheng J; Department of Radiology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, Shaanxi, China.; Shaanxi Engineering Research Center of Computational Imaging and Medical Intelligence, Xi'an, Shaanxi, China., Li B; Department of Head-Neck-Otolaryngology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, Shaanxi, China., Zhou X; Department of Pediatrics, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, Shaanxi, China., Salvi R; Center for Hearing and Deafness, University at Buffalo, Buffalo, New York, USA., Yang J; Department of Radiology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, Shaanxi, China.; Shaanxi Engineering Research Center of Computational Imaging and Medical Intelligence, Xi'an, Shaanxi, China.
Jazyk: angličtina
Zdroj: Journal of magnetic resonance imaging : JMRI [J Magn Reson Imaging] 2024 Dec; Vol. 60 (6), pp. 2332-2340. Date of Electronic Publication: 2024 May 22.
DOI: 10.1002/jmri.29450
Abstrakt: Background: Neonates with immature auditory function (eg, weak/absent middle ear muscle reflex) could conceivably be vulnerable to noise-induced hearing loss; however, it is unclear if neonates show evidence of hearing loss following MRI acoustic noise exposure.
Purpose: To explore the auditory effects of MRI acoustic noise in neonates.
Study Type: Prospective.
Subjects: Two independent cohorts of neonates (N = 19 and N = 18; mean gestational-age, 38.75 ± 2.18 and 39.01 ± 1.83 weeks).
Field Strength/sequence: T1-weighted three-dimensional gradient-echo sequence, T2-weighted fast spin-echo sequence, single-shot echo-planar imaging-based diffusion-tensor imaging, single-shot echo-planar imaging-based diffusion-kurtosis imaging and T2-weighted fluid-attenuated inversion recovery sequence at 3.0 T.
Assessment: All neonates wore ear protection during scan protocols lasted ~40 minutes. Equivalent sound pressure levels (SPLs) were measured for both cohorts. In cohort1, left- and right-ear auditory brainstem response (ABR) was measured before (baseline) and after (follow-up) MRI, included assessment of ABR threshold, wave I, III and V latencies and interpeak interval to determine the functional status of auditory nerve and brainstem. In cohort2, baseline and follow-up left- and right-ear distortion product otoacoustic emission (DPOAE) amplitudes were assessed at 1.2 to 7.0 kHz to determine cochlear function.
Statistical Test: Wilcoxon signed-rank or paired t-tests with Bonferroni's correction were used to compare the differences between baseline and follow-up ABR and DPOAE measures.
Results: Equivalent SPLs ranged from 103.5 to 113.6 dBA. No significant differences between baseline and follow-up were detected in left- or right-ear ABR measures (P > 0.999, Bonferroni corrected) in cohort1, or in DPOAE levels at 1.2 to 7.0 kHz in cohort2 (all P > 0.999 Bonferroni corrected except for left-ear levels at 3.5 and 7.0 kHz with corrected P = 0.138 and P = 0.533).
Data Conclusion: A single 40-minute 3-T MRI with equivalent SPLs of 103.5-113.6 dBA did not result in significant transient disruption of auditory function, as measured by ABR and DPOAE, in neonates with adequate hearing protection.
Evidence Level: 2.
Technical Efficacy: Stage 5.
(© 2024 International Society for Magnetic Resonance in Medicine.)
Databáze: MEDLINE