Evaluation of the ELITe InGenius and Bordetella ELITe MGB Kit for the detection and identification of B. pertussis, B. parapertussis and B. holmesii.

Autor: Martini H; Department of Microbiology, National Reference Centre for Bordetella pertussis, Vrije Universiteit Brussel (VUB), Universitair Ziekenhuis Brussel (UZ Brussel), Laarbeeklaan 101, Brussels 1090, Belgium. Electronic address: helena.martini@uzbrussel.be., Soetens O; Department of Microbiology, National Reference Centre for Bordetella pertussis, Vrije Universiteit Brussel (VUB), Universitair Ziekenhuis Brussel (UZ Brussel), Laarbeeklaan 101, Brussels 1090, Belgium., Demuyser T; Department of Microbiology, University Hospital Antwerp, Edegem, Belgium., Van Honacker E; Department of Microbiology, National Reference Centre for Bordetella pertussis, Vrije Universiteit Brussel (VUB), Universitair Ziekenhuis Brussel (UZ Brussel), Laarbeeklaan 101, Brussels 1090, Belgium., Piérard D; Department of Microbiology, National Reference Centre for Bordetella pertussis, Vrije Universiteit Brussel (VUB), Universitair Ziekenhuis Brussel (UZ Brussel), Laarbeeklaan 101, Brussels 1090, Belgium., Wybo I; Department of Microbiology, National Reference Centre for Bordetella pertussis, Vrije Universiteit Brussel (VUB), Universitair Ziekenhuis Brussel (UZ Brussel), Laarbeeklaan 101, Brussels 1090, Belgium.
Jazyk: angličtina
Zdroj: Diagnostic microbiology and infectious disease [Diagn Microbiol Infect Dis] 2024 Jul; Vol. 109 (3), pp. 116348. Date of Electronic Publication: 2024 May 11.
DOI: 10.1016/j.diagmicrobio.2024.116348
Abstrakt: Bordetella pertussis is the causative pathogen of whooping cough or pertussis, a contagious respiratory disease. Aside from serodiagnosis, laboratory confirmation of pertussis is done through PCR, as B. pertussis is difficult to culture. The ELITe InGenius instrument (ELITechGroup, France) with accompanying Bordetella ELITe MGB Kit was evaluated against a laboratory-developed assay. Both assays combine two screening (IS481, IS1001) and two confirmation targets (recA, ptxA-Pr or IS1002) for optimal sensitivity and specificity. The company's stated claims on sensitivity and reproducibility were confirmed. Accuracy testing showed full concordance between both assays for the screening targets. Minor discrepancies were seen for the B. pertussis confirmation target. Some cross-reactivity with other Bordetella species was observed for the IS481-target, however, none of these were confirmed in the ptxA-Pr target. These results show the suitability of the Bordetella ELITe MGB Kit for the detection and differentiation of B. pertussis, B. parapertussis and B. holmesii.
Competing Interests: Declaration of competing interest The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: Helena Martini reports equipment, drugs, or supplies was provided by ELITechGroup BV. If there are other authors, they declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
(Copyright © 2024. Published by Elsevier Inc.)