Autor: |
Villarreal CX, Shen X, Alhulail AA, Buffo NM, Zhou X, Ozen AC, Chiew M, Sawiak S, Emir U, Chan DD |
Jazyk: |
angličtina |
Zdroj: |
MedRxiv : the preprint server for health sciences [medRxiv] 2024 May 12. Date of Electronic Publication: 2024 May 12. |
DOI: |
10.1101/2024.05.02.24305807 |
Abstrakt: |
In this work, we demonstrate the sodium magnetic resonance imaging (MRI) capabilities of a three-dimensional (3D) dual-echo ultrashort echo time (UTE) sequence with a novel rosette petal trajectory (PETALUTE), in comparison to the 3D density-adapted (DA) radial spokes UTE sequence. We scanned five healthy subjects using a 3D dual-echo PETALUTE acquisition and two comparable implementations of 3D DA-radial spokes acquisitions, one matching the number of k-space projections (Radial-Matched Trajectories) and the other matching the total number of samples (Radial-Matched Samples) acquired in k-space. The PETALUTE acquisition enabled equivalent sodium quantification in articular cartilage volumes of interest (168.8 ± 29.9 mM) to those derived from the 3D radial acquisitions (171.62 ± 28.7 mM and 149.8 ± 22.2 mM, respectively). We achieved a shorter scan time of 2:06 for 3D PETALUTE, compared to 3:36 for 3D radial acquisitions. We also evaluated the feasibility of further acceleration of the PETALUTE sequence through retrospective compressed sensing with 2× and 4× acceleration of the first echo and showed structural similarity of 0.89 ± 0.03 and 0.87 ± 0.03 when compared to non-retrospectively accelerated reconstruction. Together, these results demonstrate improved scan time with equivalent performance of the PETALUTE sequence compared to the 3D DA-radial sequence for sodium MRI of articular cartilage. |
Databáze: |
MEDLINE |
Externí odkaz: |
|