Comparison of revision rates and epidemiological data of a single total knee arthroplasty system of different designs (cruciate retaining, posterior stabilized, mobile bearing, and fixed bearing): a meta-analysis and systematic review of clinical trials and national arthroplasty registries.
Autor: | Wagner A; Department of Orthopaedics and Trauma, Medical University of Graz, Auenbruggerplatz 5, 8036, Graz, Austria., Wittig U; Department of Orthopaedics and Trauma, Medical University of Graz, Auenbruggerplatz 5, 8036, Graz, Austria., Leitner L; Department of Orthopaedics and Trauma, Medical University of Graz, Auenbruggerplatz 5, 8036, Graz, Austria., Vielgut I; Department of Orthopaedics and Trauma, Medical University of Graz, Auenbruggerplatz 5, 8036, Graz, Austria., Hauer G; Department of Orthopaedics and Trauma, Medical University of Graz, Auenbruggerplatz 5, 8036, Graz, Austria., Ortmaier R; Ordensklinikum Linz, Barmherzige Schwestern, Seilerstätte 4, 4010, Linz, Austria., Leithner A; Department of Orthopaedics and Trauma, Medical University of Graz, Auenbruggerplatz 5, 8036, Graz, Austria., Sadoghi P; Department of Orthopaedics and Trauma, Medical University of Graz, Auenbruggerplatz 5, 8036, Graz, Austria. patricksadoghi@gmx.at. |
---|---|
Jazyk: | angličtina |
Zdroj: | Archives of orthopaedic and trauma surgery [Arch Orthop Trauma Surg] 2024 May; Vol. 144 (5), pp. 1997-2006. Date of Electronic Publication: 2024 Apr 03. |
DOI: | 10.1007/s00402-024-05286-6 |
Abstrakt: | Background: This study aimed to meta-analyze epidemiological data, revision rates, and incidences of different designs of a single Total Knee Arthroplasty System and compare these factors across different countries. Methods: A systematic review was conducted on clinical studies and arthroplasty registries of ATTUNE TKA from 1999 to 2020. The main endpoints analyzed were revision rates and epidemiological data. Results: The average age of patients was 67.8 years, with a gender distribution of 60% female and 40% male. The pooled average BMI was 29.4 kg/m 2 . Eight clinical studies showed a pooled revision rate per 100 observed CY of 0.5 (n = 1343 cases). Cumulative revision rates after 1, 3, and 5 years varied among registries, with the Swiss registry having the highest revision data (after 5 years: 6.3%) and the American registry having the lowest revision data (after 5 years: 1.7%). A comparison of the revision rates of mobile bearing and fixed bearing (41,200 cases) as well as cruciate retaining and posterior stabilized (n = 123,361 cases) showed no significant advantage in the first 5 years after implantation. Conclusion: In conclusion, pooled data from 41,200 cases of TKA with a single Total Knee Arthroplasty System in two arthroplasty registries revealed that there was no significant difference in revision rates between the mobile bearing and fixed bearing design within the first 5 years after implantation. In addition, a comparison of the revision rates in n = 123,361 cases showed no significant advantage for cruciate retaining or posterior stabilized in the first 5 years after implantation. (© 2024. The Author(s).) |
Databáze: | MEDLINE |
Externí odkaz: |