Phantom study for 90 Y liver radioembolization dosimetry with a long axial field-of-view PET/CT.

Autor: Mercolli L; Department of Nuclear Medicine, Inselspital, Bern University Hospital, University of Bern, Freiburgstrasse 18, CH-3010 Bern, Switzerland. Electronic address: lorenzo.mercolli@insel.ch., Zeimpekis K; Department of Nuclear Medicine, Inselspital, Bern University Hospital, University of Bern, Freiburgstrasse 18, CH-3010 Bern, Switzerland., Prenosil GA; Department of Nuclear Medicine, Inselspital, Bern University Hospital, University of Bern, Freiburgstrasse 18, CH-3010 Bern, Switzerland., Sari H; Advanced Clinical Imaging Technology, Siemens Healthcare AG, Lausanne, Switzerland., Rathke HG; Department of Nuclear Medicine, Inselspital, Bern University Hospital, University of Bern, Freiburgstrasse 18, CH-3010 Bern, Switzerland., Rominger A; Department of Nuclear Medicine, Inselspital, Bern University Hospital, University of Bern, Freiburgstrasse 18, CH-3010 Bern, Switzerland., Shi K; Department of Nuclear Medicine, Inselspital, Bern University Hospital, University of Bern, Freiburgstrasse 18, CH-3010 Bern, Switzerland.
Jazyk: angličtina
Zdroj: Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) [Phys Med] 2024 Feb; Vol. 118, pp. 103296. Date of Electronic Publication: 2024 Jan 27.
DOI: 10.1016/j.ejmp.2024.103296
Abstrakt: Purpose: The physical properties of yttrium-90 ( 90 Y) allow for imaging with positron emission tomography/computed tomography (PET/CT). The increased sensitivity of long axial field-of-view (LAFOV) PET/CT scanners possibly allows to overcome the small branching ratio for positron production from 90 Y decays and to improve for the post-treatment dosimetry of 90 Y of selective internal radiation therapy.
Methods: For the challenging case of an image quality body phantom, we compare a full Monte Carlo (MC) dose calculation with the results from the two commercial software packages Simplicit90Y and Hermes. The voxel dosimetry module of Hermes relies on the 90 Y images taken with a LAFOV PET/CT, while the MC and Simplicit90Y dose calculations are image independent.
Results: The resulting doses from the MC calculation and Simplicit90Y agree well within the error margins. The image-based dose calculation with Hermes, however, consistently underestimates the dose. This is due to the mismatch of the activity distribution in the PET images and the size of the volume of interest. We found that only for the smallest phantom sphere there is a statistically significant dependence of the Hermes dose on the image reconstruction parameters and scan time.
Conclusion: Our study shows that Simplicit90Y's local deposition model can provide a reliable dose estimate. On the other hand, the image based dose calculation suffers from the suboptimal reconstruction of the 90 Y distribution in small structures.
Competing Interests: Declaration of competing interest HS is a full-time employee of Siemens Healthcare AG, Switzerland. AR has received research support and speaker honoraria from Siemens. All other authors have no conflicts of interest to report.
(Copyright © 2024 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved.)
Databáze: MEDLINE