Clinical experience with the Nurotron™ cochlear implant in a Spanish-speaking postlingual population: Clinical safety and audiological results.
Autor: | Ordóñez Ordóñez LE; Clínica Universitaria Colombia (Clínica Colsanitas), Fundación Universitaria Sanitas (Unisanitas); Keralty. Grupo de investigación Anaboleas, Bogotá, Colombia; Universidad Militar Nueva Granada, Hospital Militar Central, Bogotá, Colombia. Electronic address: otorrino@unisanitas.edu.co., Angulo Martínez ES; Infecciones y Salud en el Trópico, Grupo de investigación Anaboleas, Bogotá, Colombia., Vanegas SC; Clínica Universitaria Colombia (Clínica Colsanitas), Fundación Universitaria Sanitas (Unisanitas); Keralty. Grupo de investigación Anaboleas, Bogotá, Colombia; Rehabiltación Infantil, Clínica Universitaria Colombia (Clínica Colsanitas), Fundación Universitaria Sanitas (Unisanitas); Keralty. Grupo de investigación Anaboleas, Bogotá, Colombia., Rodríguez Montoya SR; Universidad Militar Nueva Granada, Hospital Militar Central, Bogotá, Colombia; Rehabilitación Auditivo Verbal, Universidad Nacional de Colombia, Bogotá, Colombia. |
---|---|
Jazyk: | angličtina |
Zdroj: | Acta otorrinolaringologica espanola [Acta Otorrinolaringol Esp (Engl Ed)] 2024 May-Jun; Vol. 75 (3), pp. 137-147. Date of Electronic Publication: 2024 Jan 12. |
DOI: | 10.1016/j.otoeng.2024.01.002 |
Abstrakt: | Objective: to assess clinical safety and postoperative audiological outcomes in postlingual deafness Spanish speaking patients, who underwent surgery with Nurotron™ cochlear implant. Material and Methods: Retrospective descriptive case series study. We performed follow-up of complications and audiological measurements before and after cochlear implantation. Patients with bilateral severe to profound sensorineural hearing loss or patients with unilateral deafness with/without tinnitus were included. Repeated-measures within-subjects for assess pure tone thresholds and speech performance (bilingual test) with a detailed monitoring to establish security or adverse effects were performed. Analysis of variance tests, repetitive measures, were used for statistical analysis. Results: 31 patients were included, 17 (54.8%) men and 14 (45.2%) women. Mean age at the time of surgery was 49.82 ± 18.8 years. The mean follow-up of the group was 31.56 ± 9.57 months (minimum = 19.6 months and maximum = 52.50 months). As major complication one patient (3.23%) had a hard failure that required removal and re-implantation. 25.8% of the patients presented minor complications, the most frequent being vertigo/unsteadiness in 22.6%. The mean of language discrimination (free field at 65 dB SPL) was 62.19% ± 16.66; being 69.82% ± 7.35 in the group of severe to profound bilateral sensorineural hearing loss. A statistically significant reduction was observed in patients with tinnitus, assessed using the visual analogue scale, preoperative = 7.2 ± 1,6 vs postoperative (18months postoperative) = 1.7 ± 1.3 (p < .001). Conclusions: The Nurotron™ cochlear implant shows satisfactory audiological results, in accordance with what has been reported in the literature. Minor complications were similar to previous studies, but the percentage of hard failure should continue to be observed, which was higher than other reports with comparable follow-up. (Copyright © 2023 The Author(s). Published by Elsevier España, S.L.U. All rights reserved.) |
Databáze: | MEDLINE |
Externí odkaz: |