Word Recognition with a Cochlear Implant in Relation to Prediction and Electrode Position.

Autor: Franke-Trieger A; Department of Otorhinolaryngology, Head and Neck Surgery, Faculty of Medicine Carl Gustav Carus, Technische Universität Dresden, Fetscherstraße 74, 01307 Dresden, Germany., Lailach S; Department of Otorhinolaryngology, Head and Neck Surgery, Faculty of Medicine Carl Gustav Carus, Technische Universität Dresden, Fetscherstraße 74, 01307 Dresden, Germany., Shetty J; Department of Otorhinolaryngology, Head and Neck Surgery, Faculty of Medicine Carl Gustav Carus, Technische Universität Dresden, Fetscherstraße 74, 01307 Dresden, Germany., Murrmann K; Department of Otorhinolaryngology, Head and Neck Surgery, Faculty of Medicine Carl Gustav Carus, Technische Universität Dresden, Fetscherstraße 74, 01307 Dresden, Germany., Zahnert T; Department of Otorhinolaryngology, Head and Neck Surgery, Faculty of Medicine Carl Gustav Carus, Technische Universität Dresden, Fetscherstraße 74, 01307 Dresden, Germany., Neudert M; Department of Otorhinolaryngology, Head and Neck Surgery, Faculty of Medicine Carl Gustav Carus, Technische Universität Dresden, Fetscherstraße 74, 01307 Dresden, Germany.
Jazyk: angličtina
Zdroj: Journal of clinical medicine [J Clin Med] 2023 Dec 28; Vol. 13 (1). Date of Electronic Publication: 2023 Dec 28.
DOI: 10.3390/jcm13010183
Abstrakt: Background: the word recognition score (WRS) achieved with cochlear implants (CIs) varies widely. To account for this, a predictive model was developed based on patients' age and their pre-operative WRS. This retrospective study aimed to find out whether the insertion depth of the nucleus lateral-wall electrode arrays contributes to the deviation of the CI-achieved WRS from the predicted WRS.
Materials and Methods: patients with a pre-operative maximum WRS > 0 or a pure-tone audiogram ≥80 dB were included. The insertion depth was determined via digital volume tomography.
Results: fifty-three patients met the inclusion criteria. The median WRS achieved with the CI was 70%. The comparison of pre- and post-operative scores achieved with a hearing aid and a CI respectively in the aided condition showed a median improvement of 65 percentage points (pp). A total of 90% of the patients improved by at least 20 pp. The majority of patients reached or exceeded the prediction, with a median absolute error of 11 pp. No significant correlation was found between the deviation from the predicted WRS and the insertion depth.
Conclusions: our data support a previously published model for the prediction of the WRS after cochlear implantation. For the lateral-wall electrode arrays evaluated, the insertion depth did not influence the WRS with a CI.
Databáze: MEDLINE
Nepřihlášeným uživatelům se plný text nezobrazuje